  Datasheet File OCR Text:
  MINT1150 150w single output medical series www.slpower.com MINT1150 rev 1.4, 13-nov-12 1 features ? 2? x 4? x 1.3? package ? for 1u applications ? 150w w/air, 100w convection cooled ? universal input 90-264 vac ? power fail/output good signal ? approved to csa/en/iec/ul60601-1, 3 rd edition ? 2 x mopp input to output isolation ? 3 year warranty ? rohs compliant description the MINT1150 family is a high power density for a power supply in a 2? x 4? size. approved to en/iec/ul60601-1, 3rd edition, with isolation levels which satisfy the 2 mopp requirements. the MINT1150 family is ideal for portable medical devices, and many other applications where medical certifications, power density and cost are critical. the MINT1150 operates at universal input range of 90 to 264vac and wide temperature range -10c to +70c, delivering full rated output power up to +50c. in addition, these models feature power fail and dc ok signals. model selection model output current ripple & total ovp number volts w/200lfm air convection* noise** regulation threshold** MINT1150a1206k01 12v 12.5a 8.33a 0.5%rms, 1.2% pk-pk ? 5% 14.0 ? 1.1v MINT1150a1506k01 15v 10.0a 6.67a 0.5%rms, 1.0% pk-pk ? 5% 18.0 1.5v MINT1150a2406k01 24v 6.25a 4.17a 0.5%rms, 1.0% pk-pk ? 5% 28.0 2.5v MINT1150a4806k01 48v 3.13a 2.08a 0.5%rms, 1.0% pk-pk ? 5% 55.0 4.0v MINT1150a5606k01 56v 2.68a 1.79a 0.5%rms, 1.0% pk-pk ? 5% <59.9v notes: * maximum output power is 95 watts for input voltage of 90-105vac at 50c convection. for input voltage of 105vac or more, the total power is 100 watts at 50c convection. * * m e a s u r e d w i t h n o i s e p r o b e d i r e c t l y a c r o s s o u t p u t t e r m i n a l s , a n d l o a d t e r m i n a t e d w i t h 0 . 1 f c e r a m i c a n d 1 0 f l o w e s r c a p a c i t o r s . general specifications ac input 100-240vac, 10%, 47-63hz, 1 ? 120?370vdc turn on time less than 2 sec. @115vac (inversely proportional to input voltage and thermistor temperature) input current 115vac: 2a, 230vac: 1a hold-up time >12ms at full load, 120vac inrush current 264vac, cold start: will not exceed 50a signals ac power fail, dc ok
MINT1150 150w single output medical series www.slpower.com MINT1150 rev 1.4, 13-nov-12 2 general specifications (continued) input fuses f1, f2: 4a, 250vac fuses provided on all models overload protection hiccup mode earth leakage current < 3 0 0 a @ 2 6 4 v a c , 6 0 h z , n c short circuit protection provided - no damage will occur if the output is shorted. hiccup mode efficiency 89% typical at 115vac overvoltage protection ovp firing reduces output voltage to <50% of nominal in <50ms. see chart for trip range. output power 150w continuous with 200 lfm airflow, 100w convection cooled ? see chart for specific voltage model ratings. switching frequency pfc: variable 30-400khz. main converter: variable 35-180khz, 65-70khz at full load. transient response 50% load step. ? i/ ? t : < 0 . 2 a / s max volt deviation = 3% isolation input-output: 4000vac, 2 x mopp input-ground: 1800vac, 1 x mopp output-ground: 500vac ripple and noise see chart operating temperature -10c to +70c output voltage see chart temperature derating derate output power linearly above 50c to 50% at 70c voltage adjustability +/- 5% from nominal storage temperature -40c to +85c minimum load not required altitude operating: -500 to 10,000 ft non-operating: -500 to 40,000 ft. total regulation +/- 5%. see chart relative humidity 5% to 95%, non-condensing vibration operating: 0.003g/hz, 1.5grms overall, 3 axes, 10 min/axis non-operating: 0.026g2/hz, 5.0grms overall, 3 axes, 1 hr/axis shock operating: half-sine, 20gpk, 10ms, 3 axes, 6 shocks total non-operating: half-sine, 40 gpk, 10 ms, 3 axes, 6 shocks total dimensions w: 2.0? x l: 4.0? x h: 1.3? safety standards en/csa/ul/iec60601-1, 3 rd edition weight 183g mtbf 640,000 hours at 100w convection, 1,500,000 hours at 150w with 200lfm air auxiliary signals ac power fail: during normal operations, stays high. power fail: goes low with 5 ms warning before loss of output power due to ac failure dc ok: open collector logic signal goes and stays high 100ms to 500ms after main output reaches regulation.
MINT1150 150w single output medical series www.slpower.com MINT1150 rev 1.4, 13-nov-12 3 emi/emc compliance conducted emissions en55011/22 class b, fcc part 15, subpart b, class b radiated emissions en55011/22 class a, fcc part 15, subpart b, class a w/6db margin static discharge immunity en61000-4-2, criteria a, 6kv contact discharge, 8kv air discharge radiated rf immunity en61000-4-3, 3v/m. criteria a eft/burst immunity en61000-4-4, 2kv/5khz, criteria a line surge immunity en61000-4-5, 1kv differential, 2kv common-mode, criteria a conducted rf immunity en61000-4-6, 3vrms, criteria a power frequency magnetic field immunity en61000-4-8, 3a/m, criteria a voltage dip immunity en61000-4-11, 0% vin, 0.5cycle; 40% vin, 5 cycles; 70% vin, 25 cycles; criteria a line harmonic emissions en61000-3-2, class a, b, c, & d flicker test en61000-3-3, complies (dmax<6%) mechanical drawing
MINT1150 150w single output medical series www.slpower.com MINT1150 rev 1.4, 13-nov-12 4 connector information input connector j100 ground j101 dc output connector j200 signal connector j300 pin 1) ac line pin 2) empty pin 3) ac neutral 0.187? faston tab pin 1) +vout pin 4) -vout pin 2) +vout pin 5) -vout pin 3) +vout pin 6) -vout pin 1) pf/dc ok pin 2) common mating connector: molex 09-50-3031 pins= 08-52-0072 mating connector: molex 01-90020005 mating connector: amp 640250-6 pins = 640252-1 mating connector: molex 1375820-2 pins = 1375819 power fail/dc ok signals ? j300 ac p ower f ail s ignal dc ok s ignal isolation specifications parameter conditions/description min nom max units insulation safety rating input/ground input/output output/ground 2 mopp 1 mopp n/a electric strength test voltage input/ground input/output output/ground 1800 4000 500 - - vac vac vac j300 ? pin 2 j300?pin 1
MINT1150 150w single output medical series www.slpower.com MINT1150 rev 1.4, 13-nov-12 5 input specifications all specifications apply over specified input voltage, output load, and temperature range, unless otherwise noted. parameter conditions/description min nom max units input voltage 90 115/230 264 vac turn-on input voltage ramping up 82.7 vac turn-off input voltage ramping down 67.0 vac input frequency 47 50/60 63 hz inrush current limitation 264vac, cold start - - 50 a power factor v i nom , i o nom 0.9 - - efficiency v i nom , i o nom MINT1150a1206k01 MINT1150a2406k01 MINT1150a4806k01 MINT1150a5606k01 - 89% - % protection parameter conditions/description min nom max units input fuse not user accessible input transient protection 2kv(cm) and 1kv(dm) surge 2 kv (cm) output no-load normal short circuit hiccup mode overload hiccup mode overvoltage protection latch style see models chart for trip ranges over temperature protection automatic power shutdown at t c =155c characteristic curves output vs. temperature 100w convection cooled and 150w continuous with 200 lfm airflow, derating output power to 50% at 70 ? c.
MINT1150 150w single output medical series www.slpower.com MINT1150 rev 1.4, 13-nov-12 6 efficiency vs. loading the high efficiency is achieved by using llc technology, pfc topology minimizing switching losses. synchronous mosfet or schottky diode is used as rectifier in MINT1150 family. ripple & noise to verify that the output ripple and noise does not exceed the level specified in the product specification, measured using a scope probe socket with 0.1uf ceramic and a 10uf electrolytic capacitor connected in parallel across it, 20mhz bw. 24v out , n o l oad , 90v ac , 60h z 24v out , f ull l oad , 90v ac , 60h z 24v out , n o l oad , 264v ac , 50h z 24v out , f ull l oad , 264v ac , 50h z
MINT1150 150w single output medical series www.slpower.com MINT1150 rev 1.4, 13-nov-12 7 output transient response 5 0 % l o a d s t e p w i t h i n t h e r e g u l a t i o n l i m i t s o f m i n i m u m a n d m a x i m u m l o a d , d i / d t < 0 . 2 a / s e c . r e c o v e r y t i m e n o t s p e c i f i e d a s t h e r e i s n o l a p s i n r e g u l a t i o n with a 50% load step. maximum voltage deviation is 3%, this test is performed on the main output only. 24v out , 120v ac , 25% to 75% l oad s tep 24v out , 240v ac , 25% to 75% l oad s tep output overload characteristic supply shall protect itself against overload conditions. the power supply shall recover from overload conditions without operator intervention. 24v out , 90v ac 24v out , 264v ac short circuit protection supply shall protect itself against short circuit conditions. no damage will occur if the output is shorted. 24v out , 90v ac 24v out , 264v ac
MINT1150 150w single output medical series www.slpower.com MINT1150 rev 1.4, 13-nov-12 8 overvoltage protection ovp firing reduces output voltage to <50% of nominal in <50ms. see models chart for trip ranges. 24v out , n o l oad turn on time 90v ac , f ull l oad 264v ac , f ull l oad hold up time 120v ac , f ull l oad data sheet ? 2012 sl power electronics corp. the information and specifications contained herein are believed to be correct at the time of publication. however, sl power accepts no responsibility for consequences arising from reproduction errors or inaccuracies. specifications are subject to change without notice.


▲Up To Search▲   

 
Price & Availability of MINT1150

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X